Final Decision to Certify Hazardous Waste Environmental Technology Millipore, Inc. Enviro-Gard® An enzyme immunoassay for fast, semiquantitative field measurements of polychlorinated biphenyls in soil The following is excerpted from: CALIFORNIA REGULATORY NOTICE REGISTER The California Environmental Protection Agency, Department of Toxic Substances Control (Department) has made a final decision to certify the following company's hazardous waste environmental technology listed below:
A new law effective January 1, 1994 (Chapter 412, Statutes of 1993, Section 25200.1.5., Health and Safety Code, enacted by Assembly Bill 2060, Weggeland), authorizes the Department to certify the performance of hazardous waste environmental technologies. Only technologies that are determined to not pose a significant potential hazard to the public health and safety or to the environment when used under specified operating conditions and which can be operated without specialized training and with minimal maintenance may be certified. Incineration technologies are explicitly excluded from the certification program. The purpose of the certification program is to provide an in-depth, independent review of technologies at the manufacturers' level to facilitate regulatory and end-user acceptance and to promote and foster growth of California's environmental technology industry. The Department makes no express or implied warranties as to the performance of the manufacturer's product or equipment. The end-user is solely responsible for complying with the applicable federal, state, and local regulatory requirements. Certification does not limit the Department's authority to require additional measures for protection of public health and the environment. By accepting certification, the manufacturer assumes, for the duration of certification, responsibility for maintaining the quality of the manufactured equipment and materials at a level equal to or better than was provided to obtain certification and agrees to be subject to quality monitoring by the Department as required by the statute under which certification is granted. DTSC's notice to certify was published in the California Regulatory Notice Register Volume 94, No. 27-Z. The DTSC's final certification shall be effective from August 8, 1994 to August 8, 1997. Additional information supporting DTSC's final certification decisions is available from:
A description of the technology to be certified, the final certification statement and the certification limitations for the technology follows: Millipore Corporation Technology Description The technology is an enzyme immunoassay for fast, semiquantitative field measurements of polychlorinated biphenyls (PCBs, as Aroclors) in soil. The reaction is performed on a methanol extract of a small sample of the soil. The system is designed to give decisions on Aroclor content at 5, 10, or 50 ppm (Aroclors 1242 through 1260). The manufacturer provides testing reagents and utensils, a user manual, material data safety sheets, and instruction in the use of the testing system. Certification Statement Under the authority of Section 25200.1.5 of the California Health and Safety Code, the Department hereby certifies the EnviroGard™ Immunoassay for PCBs in Soil manufactured by Millipore Corporation as a Measurement Technology. The test kit consists of a semiquantitative immunoassay system for the detection of PCBs, as Aroclor mixtures. Provided that the immunoassay is used properly, detection of Aroclor 1242 through 1260 is at the 5, 10, and 50 ppm levels. For other Aroclors, higher target levels apply. The assay is not suitable for Aroclors 1221 and 1232. The technology is applicable to testing of soil or soil-like material which is not highly contaminated with oil or other non-aqueous phase liquids (NAPLs). The efficiency of the test depends on the extraction of PCBs from the test soil into a methanol solution. The efficiency of extraction is reduced in excessively moist soils and may be affected by high clay or organic content of the soil. It is recognized that the calibration is biased so as to minimize the possibility of false negative results; other limitations are set forth below. The Department's findings are described in greater detail in an evaluation report. Limitations of Certification The Department makes no express or implied warranties as to the performance of the manufacturer's product or equipment. The Department has not conducted any bench or field tests to confirm the manufacturer's performance data. Nor does the Department warrant that the manufacturer's product or equipment is free from any defects in workmanship or material caused by negligence, misuse, accident, or other causes. The Department believes, however, that the manufacturer's product or equipment can achieve performance levels set out in this certification. Said belief is based on a review of the data submitted by the manufacturer and other information, provided that the product is used in accordance with the manufacturer's specifications. Specific Conditions
Basis for Certification The following documents were available for this evaluation:
Recommended Applications of the Test System The immunoassay is for the semiquantitative determination of PCBs in terms of a commercial mixtures of PCBs (Aroclor 1248) that has been used in the design and calibration of the assay. Conversion factors are applied to the results for Arclors to which the assay responds differently. Unknown PCB mixtures need to be characterized by a reference method and a conversion factor needs to be determined which reflects the immunoassay response of that mixture relative to the response of Aroclor 1248. Without such an adjustment, results can be either high or low, depending on the affinity for the assay's antibodies of an unknown PCB mixture. A semiquantitative determination will provide a response, interpreted as either positive or negative, at one or several predetermined detection or target levels. Target levels are usually chosen to have relevance to a specific situation. A comprehensive process of developing data quality objectives (DQO) was published by U.S. EPA under the U.S. Superfund Program. It provides guidance for analytical method QA/QC as applied to field investigations for PCB-contaminated soils. The process is intended for site-specific sampling plans. Here the immunoassay would generally qualify as a Level 2 (field analysis) method, subject to confirmation by a Level 3 method (identification and quantification, i.e., EPA Methods 8080 or 8081) applied predominantly to positive results. We recommend that minimum quality control should include method blanks and duplicates at 5 percent, or one per batch or per matrix, whichever is the more frequent, in addition to the samples required for confirmation. The use of proficiency evaluation and spiked samples should depend on project-specific needs. We recommend gas-chromatographic U.S. EPA Methods 8080 or 8081 for establishing or confirming the types and concentrations of Aroclor(s). "Screening" and Preliminary Site Investigations--The immunoassay can assist in preliminary site investigations ("Phase I"), if there are compelling historical data to indicate the presence of PCBs. If used on samples of largely unknown composition, without prior characterization by an approved, fully qualitative and quantitative laboratory method, confirmatory analysis is needed for every positive immunoassay result. Negative results need to be evaluated in light of the known specificity of the assay and the nature of the known contaminant(s). The assay design is such that matrix effects and/or interferences will tend to yield positive results which may not be due to the presence of the contaminant. A margin of error (above the stated detection level) is in order for those PCBs that may show a lower response than those for which the assay has been calibrated. In the absence of other regulations and guidelines, we recommend that assay results be confirmed in the following manner:
If the appropriate protocols are followed, the immunoassay can be used to great advantage to classify contaminated soils as to low, medium, or high contamination and to determine which samples would provide the most information from laboratory analysis. Site Investigations and Remedial Actions--Here the testing is expected to proceed under a site-specific Quality Assurance Project Plan (QAPP). Immunoassay and other field measurements will be "bracketed" in time and space by qualitative and fully quantitative analyses. Generally, a site is first characterized by the use of approved, fully qualitative and quantitative analytical methods as to the nature and level of contamination in key sampling locations and as to the presence of substances that may interfere with the use of the immunoassay. After such initial characterization, the immunoassay can be used in the comprehensive mapping of the site with respect to identified contaminant(s) to which the immunoassay responds. The percentage of samples that would be confirmed by another approved, fully quantitative method would be as stipulated in the QAPP; the project manager could call for additional confirmatory testing if such a need is indicated in the course of the investigation. During site cleanup, the QAPP would provide for use of the immunoassay to monitor progress. Confirmatory laboratory testing would occur before a decision on site closure is made. Regulatory Implications As of April 6, 1993, this immunoassay has been accepted for use in Draft Method 4020 by the U.S. EPA Office of Solid Waste (SW-846 Collection of Methods, Revision Draft 1, October 1992). The Department's Certification is based on the technology's performance and by itself does not change the regulatory status of PCB testing; it should, however, facilitate and encourage the acceptance of this technology where a project's data quality objectives can be met by its use. To this end, the Department's findings should contribute to a consideration of this technology in regulated activities, depending on each regulated program's objectives and constraints. State certification does not imply certification by the U.S. Government for use at federal superfund sites and other facilities under the jurisdiction of the U.S. Government for which state authorization for administrative oversight has not been granted. Under state implementation of the U.S. Resource Conservation and Recovery Act (RCRA), Treatment, Storage, and Disposal Facilities may contact state permitting agencies for use of the immunoassay for operational monitoring as part of a Waste Analysis Plan (WAP). This Certification is issued as part of a pilot project to expedite the California Environmental Technology Certification Program. As a result, this Certification is subject to the conditions set out in the regulations to-be-developed, such as the duration of the Certification, the continued monitoring and oversight requirements, and the procedures for certification amendments, including decertification. By accepting this Certificate, the manufacturer assumes, for the duration of the Certification, responsibility for maintaining the quality of the manufactured materials and equipment at a level equal or better than was provided to obtain this Certification and agrees to be subject to quality monitoring by the Department as required by the law under which this Certification is granted. For more information, contact us at: